Symptoms
When starting the MiraQ the following error message appear:
Solution
Check system DHR in order to determine if system should have DICOM support or not.
DICOM support
Check the system report for the following entry:
If system should have DICOM and the line "DICOM API initialized is set to NO you need to import a valid DICOM license file and an updated USB license file
No DICOM support
Check the system report for the following entry:
If the system should not have DICOM and the line "DICOM API initialized is set to YES a possible cause can be if settings.mdb has been altered in some way setting HasDicomSupport = True. The system will then check if a valid DICOM license is absent. If this is the case here the problem can be solved by importing "NoDicomSettings.xml"